Chưa có sản phẩm trong giỏ hàng!
Cần tư vấn thêm? Hotline/Zalo: 0905 412 411 - 0938 861 947
Giá cạnh tranh
Hàng chính hãng 100%
Giao hàng nhanh
Tên sản phẩm: Water Baths BM510
NSX: YAMATO
THÔNG TIN CHI TIẾT
Bếp cách thủy Water Baths BM510 YAMATO
BM510 YAMATO
Easy to remove, clean the water tank and change water
SERIES BM YAMATO
Model |
100 | 200 | 400 | 500 | 510 |
Operating temperature range |
Room temp. +5deg.C to 95 deg.C |
Room temp. +5deg.C to 90 deg.C |
Accuracy of Temp. Adjustment |
±2.0deg.C | ±1.0deg.C | ±1.5deg.C (When stirring) |
Bath Capacity |
3.7 L | 7.0 L | 4L |
Specifications
BM500 | BM510 | BM100 | BM200 | BM400 |
Performance |
Operating temperature range*1 | Room temp. +5deg.C to 90 deg.C |
Room temp. +5deg.C to 95 deg.C |
Temp adjustment |
±1.5deg.C (When stirring) | ±2deg.C (at 60deg.C) | ±1deg.C (at 60deg.C) |
Components |
Control system | PlD control by microprocessor | ON/OF control | PlD control by microprocessor |
Set-up system | Digital setting by key | Analog set-up | Digital setting by key |
Extra function | Keylog function, SE signal transmission and reception, power failure guarantee operation, calibration offset | No extra function | Temp. preset function (preset/select one temp.) |
Sensor | K thermocouple | Liquid expansion type | Pt 100Ω |
Heater | 1kW (at 100V) 1.44kW (at 120V) | 1kW (at 200V) 1.44kW (at 240V) |
SUS316 pipe heater, 500W |
SUS316 pipe heater, 1000W |
SUS316 pipe heater, 1000W |
Safe Device |
Safety countermeasures |
Protection cover for bath. Self-diagnostic functions (Automatic overheat prevention, Sensor trouble, Triac short circuit, heater burn out, main relay failure), socket breaker, thermal fuse |
Protection from over current (fuse), Protection cover for bath, Thermal fuse |
Self-diagnostic functions (Automatic overheat prevention, Sensor trouble, Triac short circuit, heater burn out, main relay failure), socket breaker, thermal fuse |
Interlock control function |
Automatic control signal receive by RE600/800 controller, Error report to RE600/800 | _ |
Standard |
Bath Capacity/Material | 4L | 4.7 L | 7.0L |
Water tank | ø 240 x bottom 165 X depth 122mm | ø 200×Depth 120mm | ø250×Depth 150mm |
External dimensions (W×D×Hmm)*3 |
340 x 349 x 231 | 240×300×150 | 310×360×230 |
Power source 50/60Hz |
AC100V- 120V 10.5 – 12.5A (15A) |
AC200-240V 5.5 – 6.5A (10A) | AC100V 6A with paddle switch |
AC100V 12A | AC100V 12A (20A) |
Weight | Approx. 5.5kg | Approx. 3.5kg | Approx. 6.0kg | Approx. 7.0kg |
Accessories |
Bar thermometer (10 to 110deg.C) with immersion line |
*1. In case of unloaded operation of bath only. The maximum temperature varies on circumstances and
operational conditions.
*2. The performance shown here is the value under the condition of the ambient temperature of 23deg.C±5deg.C,
humidity of 65%RH±20%.
*3. The projection is not included for Internal dimensions and External dimensions.